The most common way for people with diabetes and their healthcare team to assess glycaemic control is through keeping HbA1c and fingerprick test diaries. However, this information alone is sometimes not enough to adjust treatment effectively.
Situations requiring detailed information about blood glucose fluctuations that only continuous glucose monitoring (CGM) can provide include: when adjusting therapy; quantifying the response in a trial of a diabetes therapy; assessing the impact of lifestyle modifications on glycaemic control; monitoring conditions where tighter control without hypoglycaemia is sought (for example, gestational diabetes); diagnosing and then preventing hypoglycemia (for example, during sleep); and diagnosing and preventing postprandial hypoglycaemia.
The most important use of CGM is to facilitate adjustments in therapy to improve glycaemic control and/or quality of life (Sabbah et al, 2000; Kaufman et al, 2001). When deeper insight into glycaemic patterns is needed, CGM can help supplement HbA1c and fingerprick values, thus providing a complete picture of a person’s metabolic state. CGM is also a powerful tool for education and motivation.
Using continuous glucose monitoring
CGM devices work by measuring the glucose in the interstitial fluid rather than in capillary blood. Interstitial fluid is the main component of the extracellular fluid, and is found in the interstitial spaces. On average, a person has about 11 litres of interstitial fluid, which provide the cells of the body with nutrients and a means of waste removal.
When using CGM, a sensor is inserted into subcutaneous tissue, usually on the abdomen or upper arm, which feeds the information to a receiver using either a transmitter worn on the skin or directly via a cable. Depending on which device is used, the receiver may display the reading every few minutes to the user, and/or store the information for downloading after the sensor is removed. Once the information has been downloaded, computer software is used to analyse the readings.
Advantages and disadvantages of CGM
Advantages
Most CGM systems provide up to 288 glucose readings per day (average of one test every 5 minutes), providing a level of detail about a person’s glycaemic pattern that would otherwise be unattainable. CGM allows early detection of glucose trends and identification of previously undetected glucose patterns. For example, CGM would allow detection of unrecognised nocturnal hypoglycaemia. In this case, therapy or lifestyle can then be adjusted for improved glycaemic control (Kaufman et al, 2001; Pitzer et al, 2001).
CGM allows examination of how blood glucose levels react to insulin, exercise, food, and other factors. These data can be useful for setting correct insulin dosage for food intake and correction of hyperglycemia (Sabbah et al, 2000). Monitors may also be equipped with alarms to alert people of hyper- or hypoglycemia so that they can take corrective action even in cases where they do not yet feel any symptoms.
CGM detects four times more high and low blood glucose levels than fingerprick tests alone, and studies have demonstrated that people with continuous sensors experience less hyperglycemia (Schiaffini et al, 2002; Weintrob et al, 2004; Garg et al, 2004; 2006), and can generally maintain lower HbA1c levels compared with fingerprick test management alone (Bode et al, 1999; Kaufman et al, 2001; Pitzer et al, 2001; Ludvigsson and Hanas, 2003; Deiss et al, 2006; Garg and Jovanovic, 2006; Mastrototaro et al, 2006).
Some CGM systems are available with real-time values displayed for the person to see, or with stored values only – which means the results can only be viewed after removal of the sensor and downloading the results onto a computer. There are benefits to both systems. Real-time systems can show the person, at the time, the effects of insulin doses and lifestyle changes, whereas monitors that store the readings are useful if trends are to be viewed over a period of days without the person making any corrections.
Disadvantages
Glucose levels in interstitial fluid lag behind blood glucose values by about 5 minutes (Steil et al, 2003; Wentholt et al, 2005; Wilhelm et al, 2006), therefore people are required to undertake fingerprick measurements for CGM system calibration and are often advised to use fingerprick measurements to confirm hypo- or hyperglycaemic status before taking corrective action.
CGM systems require up to four fingerprick blood glucose measurements per day for calibration. The ideal time to calibrate is either after fasting or at least 3 hours postprandially, but not immediately after exercise or when the blood glucose level is likely to be rising or falling due to the lag-time with interstitial fluid. Without such calibration, continuous readings may be incorrect. People with diabetes require a thorough training programme to calibrate and operate a CGM system.
Blood glucose levels, when changing rapidly, may read as being in a normal range on a CGM system while in reality the person is already experiencing symptoms of an out-of-range blood glucose value and may require treatment. People using CGM are therefore advised to consider both the absolute value of the blood glucose level given by the system as well as any trend in the blood glucose levels. For example, a person using CGM with a blood glucose of 5 mmol/L might take no action if their blood glucose has been consistent for several readings, while a person with the same blood glucose level but whose blood glucose has been dropping in a short period of time might be advised to perform a fingerprick test to check for hypoglycemia.
Audit
Aim
An audit was undertaken at the author’s institution to determine whether a 72-hour glucose profile can provide information to improve diabetes control and to assess the long-term effects on HbA1c levels. Ethical approval was not needed for this study as it was designed and conducted to produce information to inform delivery of best care and is therefore classed as a clinical audit (National Patient Safety Agency, 2009).
The study recruited 12 people (seven women and five men) with type 1 diabetes being treated with multidose insulin therapy. Mean age was 57 years (age range 27–80 years). Inclusion criteria were:
- Age 18 years or older.
- Type 1 diabetes.
- Currently on a basal–bolus regimen.
- Those whose diabetes control did not appear to be improving despite regular review and ongoing management under the outpatient care of the author’s diabetes centre.
Method
People with diabetes attending the diabetes clinic who fulfilled the inclusion criteria were offered CGM to determine if any additional information would lead towards improvements in their diabetes control. Those who agreed to undergo CGM completed a food and insulin diary for the duration of the monitoring and followed a care pathway, which comprised:
- Training on the use of the CGM and insertion of the sensor.
- Removing the device either at the clinic or by themselves at home. They were advised to bring their food and insulin diary with them when they returned the monitor to the clinic.
- Attending a session with the DSN and diabetes specialist dietitian to discuss their results and compare them with their diary. Recommendations would then be made for lifestyle and nutrition changes.
- Completing a short questionnaire on their experience of using CGM.
- Follow-up at 3, 12, 24 and 36 months.
Results
HbA1c results at 3, 12, 24 and 36 months post-CGM were noted and compared with their initial pre-CGM result. Before CGM, the mean HbA1c level of the group was 8.8+1.4% (73±15.3 mmol/mol). The results of the monitoring showed previously unrecognised blood glucose excursions in all participants. Seven suffered unrecognised nocturnal hypoglycaemia, with an average of two asymptomatic nocturnal and early morning hypoglycaemic events per person (ranging between one and three episodes). Prolonged periods of hyperglycaemia (blood glucose levels >14 mmol/L for 5 hours) were also recorded in three people.
At 3-month follow-up, mean HbA1c levels had reduced from 8.8% (73 mmol/mol) to 8.2% (66 mmol/mol; P=0.028). This improvement was attributed to lifestyle and diet modification rather than changes in therapy. Mean HbA1c levels at 12, 24 and 36 months were 8.2% (66 mmol/mol; P=0.009), 8.0% (64 mmol/mol; P=0.009) and 7.5% (58 mmol/mol; P=0.004), respectively.
The lack of reduction in mean HbA1c level after 12 months compared with the mean level at 3 months was a result of one person’s HbA1c increase from 6.7% (50 mmol/mol) to 8.2% (66 mmol/mol), which had a significant effect on the mean HbA1c level of the group. The reason for this increase was unclear, but at 36 months this person’s HbA1c level had come back down to 5.9% (41 mmol/mol).
By study end, 11 people had showed an improvement in their HbA1c level, while one person remained the same. Of the 11 who showed an improvement, four improved by up to 1 percentage point, five had improved by between 1.1 and 2 percentage points, and two people improved by over 2 percentage points.
Participants’ experience of CGM
During the initial 3-month assessment, participants completed a short questionnaire on their experience using the CGM system. Generally, they were positive about the experience. The results are shown in Table 1.
Discussion
CGM was performed for 72 hours and results were discussed with the participants. Discussion took place in a joint consultation with a DSN and a diabetes specialist dietitian. The participants completed a diet and lifestyle diary while on CGM, and this was used during the consultation to see how diet and other lifestyle issues such as exercise affected their glucose levels. Using this information, insulin or lifestyle adjustments were recommended.
It was found that CGM offered advantages over intermittent glucose monitoring when glycaemic patterns were poorly understood. The information about direction, magnitude, duration, frequency, and causes of fluctuations in blood glucose levels that can be obtained by CGM is simply not available with intermittent blood glucose monitoring; in this study this was most noticeable from the night-time glucose levels, after ingesting food, and during and after exercise or increased activity. When patterns are needed to adjust therapy or identify areas of hyper- or hypoglycaemia, CGM is useful.
A monitor that does not give real-time readings was used as this was the only one available to the team, and it was found that the information proved valuable for management decisions regarding peoples’ diabetes care. During hypoglycaemia or periods of rapid fluctuation, values provided by CGM may be inaccurate and confirmation is needed by means of a fingerprick test. Clinical outcome studies suggest that measures of mean glycaemia and hypoglycaemic burden both improve with the use of CGM (Bode et al, 1999; Tanenberg et al, 2004).
Individualised care plans were made for each person, and agreed changes made and targets aimed for.
Conclusion
CGM is often seen as expensive, and the monitors and sensors can be costly. However, CGM can greatly improve a person with diabetes’ quality of life as a result of reduced risk of hypoglycaemia, improved glycaemic control and the subsequent reduced risk of long-term vascular complications.
CGM may become a routine part of diabetes management in the future, initially for people with difficult-to-control diabetes and eventually for most people with diabetes.
Retrospective reporting, as used in this small study, will eventually give way to real-time readings, and use requiring a confirmatory fingerprick blood test will eventually give way to use without the requirement of such calibration. As methods for minimally invasive and non-invasive monitoring are developed, people with diabetes will use this technology more routinely. Data printouts from CGM will increasingly provide information for effective diabetes management in the future.
Su Down looks back on a year of change and achievement.
17 Dec 2024